#NanoVNA-H Firmware for 75 Ohm port?
#nanovna-h
Without changes in the firmware it will not work properly. Of course, if you're only interested in SWR it's probably enough, but you'll see 50 ohms instead of 75 when measuring impedance.
Від
btomek@...
· #38826
·
|
#NanoVNA-H Firmware for 75 Ohm port?
#nanovna-h
In this firmware https://github.com/DiSlord/NanoVNA-D/releases/tag/v1.2.40 is a option DISPLAY/PORT-Z where you can specify impedance, ex 75, or 100 etc. Calibrate normally before entering PORT-Z with
Від
btomek@...
· #38816
·
|
dfu_util command line for updating firmware please ?
These params work for me: for dfu file: dfu-util -a 0 -D filename.dfu for bin file: dfu-util -d 0483:df11 -a 0 -s 0x08000000:leave -D filename.bin
Від
btomek@...
· #33702
·
|
NanoVNA-App in app calibration problem (bug?)
#calibration
#nanovna-app
@Roger, thank you for these tests that confirm the problem, you did a good job! @Siegfried - after all, the whole topic is about getting an incorrect result despite calibrating the cable. Did you meas
Від
btomek@...
· #31677
·
|
NanoVNA-App in app calibration problem (bug?)
#calibration
#nanovna-app
@Brian - It may be relevant. Comparison of Open in calibrated NanoVNA-App and calibrated NanoVNA-Saver. In Saver it looks better, oscillating around 1 and in NanoVNA-App it is > 1 for all points.
Від
btomek@...
· #31635
·
|
NanoVNA-App in app calibration problem (bug?)
#calibration
#nanovna-app
@DiSlord - But that's exactly right. The measurement range is identical to the calibration range, the number of points is also the same, the calibration frequencies are equal to the measurement freque
Від
btomek@...
· #31632
·
|
NanoVNA-App in app calibration problem (bug?)
#calibration
#nanovna-app
@Brian - My s1p was generated in NanoVNA-App. Whether there is a period or a comma probably depends on operating system's regional settings. According to wikipedia https://en.wikipedia.org/wiki/Decima
Від
btomek@...
· #31598
·
|
NanoVNA-App in app calibration problem (bug?)
#calibration
#nanovna-app
Roger, thank you for reproducing the experiment. We have clarity that it is a Bug after all. I did an experiment in which I change the calibration data for a 4 specific frequencies by editing a copy o
Від
btomek@...
· #31595
·
|
NanoVNA-App in app calibration problem (bug?)
#calibration
#nanovna-app
Could someone be so kind as to reproduce this measurement at his computer, on his copy of the NanoVNA-App, with his NanoVNA and his cable of several meters? If the result is good, it means that I am d
Від
btomek@...
· #31582
·
|
NanoVNA-App in app calibration problem (bug?)
#calibration
#nanovna-app
Hello Roger and John. I did not write exactly what the s1p files are, sorry. short.s1p open.s1p = These are the cable measurement files with calibration in the NanoVNA-App application. These are the r
Від
btomek@...
· #31572
·
|
NanoVNA-App in app calibration problem (bug?)
#calibration
#nanovna-app
Hello. If I connect a cable to the NanoVNA and calibrate OSL at the end of that cable, and measure the same cable, I should get 1 small point to the right of the Smith chart when the cable is open, or
Від
btomek@...
· #31552
·
|
recall saved images NanoVNA-h4
Recal saved s1p bmp and cal data is possible in DiSlord firmware in sd card menu. I'm not sure from which version, but I think it's 1.2.00. DiSlord commited this option to open source version at Dec 2
Від
btomek@...
· #31100
·
|
Measuring FM deviation for 9600 baud packet
For 9600 transmission you need an FM modulator whose deviation directly depends on the amplitude of the signal. There can be no filtering or preemphasis/deemphasis there. In other words, you feed a si
Від
btomek@...
· #30744
·
|
Wireless control of NanoVNA
In the current firmware there is an option to control the device through the serial port, instead of the usb port. (Menu/Config/Connection/Serial and Serial Speed) I used a Serial/Bluetooth HC-05 conv
Від
btomek@...
· #30419
·
|
#crash
#crash
You cannot upload normal firmware to S-A-A V2.2 with DfuSE.In this version of NanoVNA there is a bootloader at the beginning 0x08000000. If you have uploaded firmware from this address, you have overw
Від
btomek@...
· #26309
·
|
Data Points Question
You will only have 3-4 calibration points in the 3-4MHz range, and 101 measurement points. If the error is large, you will notice ripples on the graph. I recently measured an antenna on a car using a
Від
btomek@...
· #25712
·
|
two nanoVNA linked for two port setup
How is the sweep synchronized in both of these devices? Unless it doesn't have to be synchronized, but it seems to me that it does.
Від
btomek@...
· #25068
·
|
Measuring different results
David Reichard: That's not what I thought about, maybe it's a language barrier issue, I'm using a translator.The SWR meter designed for 50 Ohms is only able to measure SWR for a 50 Ohm antenna system,
Від
btomek@...
· #25021
·
|
Measuring different results
The NanoVNA measuring bridge at the S11 connector is in equilibrium at a load of 50 Ohms, and works most accurately in this region. The further away from 50 Ohms the more error there is in the measure
Від
btomek@...
· #25012
·
|
Sweeping an Antenna from the Shack
If you calibrate at the end of the cable, there should be no difference in either case. But I would choose option 1 because the attenuation of 18 feet of LMR-100 at 145MHz is only 1.62 dB and the perf
Від
btomek@...